model-270

Model 270

Advanced diagnostic audiometer

An advanced diagnostic audiometer providing an extensive range of measurements for diagnostic audiology and hearing aid fitting.

Including air conduction (AC), bone conduction (BC) and speech testing, the Model 270 is ideal for audiology specialists and hearing health experts who operate within a clinic, workplace, or on-the-go.

model-270

Key features

  • Frequency range (AC): 125 - 8000Hz
  • Attenuation range (AC): -10 - 120dBHL
  • Air and bone conduction with masking
  • THL, MCL, ULL/UCL
  • Speech testing (live and recorded)
  • Free field audiometry (2 speakers)
  • Special tests: ABLB, Stenger, SISI
  • Third-party EMR connectivity
  • Compact, lightweight and portable

Supportive design

The Model 270 benefits from having a clear and precise display at an optimal viewing angle, a logical key layout, and durable design. Operators can access functions quickly and easily to reduce test times. The addition of silent rotary controls, also helps to minimise noise levels for more accurate results.

Special tests

The Model 270 also includes clinical tests such as Stenger, ABLB and Short Increment Sensitivity Index (SISI) ensuring that class leading patient care can be provided. Watch this 3 minute video to learn more:

Request more information

Included
  • Standard audiometric headset
  • Bone conduction headset
  • Patient response switch
  • Power supply with country adaptors
  • Audiogram cards (50)
  • Carry case
  • Manual & software (available via website download)
Optional
  • Amplivox Audiocups (noise-reducing enclosures)
  • Masking earpiece
  • Insert earphones
  • Thermal printer
  • OtoAccess® database
  • Combined microphone and monitor headset
  • Talk back microphone
  • Loudspeaker kit
  • Booth leads
  • USB cable (PC connection)
  • USB portable power bank (external)
Consumables

High-quality consumables and accessories to support patient testing programmes, reduce the risk of infection and optimise the performance and reliability of your equipment.

Specifications
Frequency range: 125 - 8000Hz
Attenuation range: -10 - 120dBHL

CPT codes
92552: Pure tone audiometry (threshold); air only
92553: Pure tone audiometry (threshold); air and bone
92557: Comprehensive audiometry threshold evaluation and speech recognition
92562: Loudness balance test, alternate binaural or monaural
92565: Stenger test, pure tone
92579: Visual reinforcement audiometry (VRA)
92582: Conditioning play audiometry

Standards
Audiometer Type 3BE/Type 2
IEC 60601-1
IEC 60601-1-2
IEC 60645-1/ANSI S3.6

Two year warranty as standard. Please note: product availability may differ depending on country regulation and registration. For further information please contact us.